Abstract
"Stress shielding" of bone around noncemented prosthetic hip stems causes long-term adaptive bone resorption which threatens the integrity of the fixation. Recently, computer-simulation models based on adaptive bone-remodeling theory in combination with finite-element methods have been developed. These models can be used to predict the extent of long-term resorption. In this paper a comparison is presented between the results of these predictions and those of precise measurements in retrieval hip-replacement specimens reported by Engh and associates (1992). It is shown that certainly the predictions and the actual clinical findings are subject to the same trends. Under certain assumptions, these results even match precisely. It is also shown that both clinical results and simulation predictions can be approximated by a simple formula, in which the amount of bone loss is related to the ratio between stem stiffness and preoperative bone density.
Similar articles
-
The relationship between stress shielding and bone resorption around total hip stems and the effects of flexible materials.Clin Orthop Relat Res. 1992 Jan;(274):124-34. Clin Orthop Relat Res. 1992. PMID: 1728998
-
Effects of material properties of femoral hip components on bone remodeling.J Orthop Res. 1992 Nov;10(6):845-53. doi: 10.1002/jor.1100100614. J Orthop Res. 1992. PMID: 1403299
-
Stemmed femoral knee prostheses: effects of prosthetic design and fixation on bone loss.Acta Orthop Scand. 2002 Dec;73(6):630-7. doi: 10.1080/000164702321039589. Acta Orthop Scand. 2002. PMID: 12553509
-
Mechanisms of bone loss associated with total hip replacement.Orthop Clin North Am. 1993 Oct;24(4):583-90. Orthop Clin North Am. 1993. PMID: 8414423 Review.
-
Validation of adaptive bone-remodeling simulation models.Stud Health Technol Inform. 1997;40:33-48. Stud Health Technol Inform. 1997. PMID: 10168881 Review.
Cited by
-
A Comprehensive Review on Manufacturing and Characterization of Polyetheretherketone Polymers for Dental Implant Applications.3D Print Addit Manuf. 2024 Aug 20;11(4):1441-1461. doi: 10.1089/3dp.2023.0064. eCollection 2024 Aug. 3D Print Addit Manuf. 2024. PMID: 39360128 Review.
-
Biomechanical evaluation of intramedullary retrograde nail and dynamic condylar screw used in unstable distal femoral fractures.Phys Eng Sci Med. 2023 Dec;46(4):1755-1764. doi: 10.1007/s13246-023-01335-3. Epub 2023 Oct 23. Phys Eng Sci Med. 2023. PMID: 37870727
-
The effect of humeral implant thickness and canal fill on interface contact and bone stresses in the proximal humerus.JSES Int. 2021 Jul 8;5(5):881-888. doi: 10.1016/j.jseint.2021.05.006. eCollection 2021 Sep. JSES Int. 2021. PMID: 34505100 Free PMC article.
-
Zeolite Socony Mobil-Five Coating on Ti-24 Nb-4 Zr-7.9 Sn Promotes Biocompatibility and Osteogenesis In Vitro and In Vivo.Biomed Res Int. 2021 Jul 29;2021:5529368. doi: 10.1155/2021/5529368. eCollection 2021. Biomed Res Int. 2021. PMID: 34368350 Free PMC article.
-