Individualized correction of neuron-specific enolase (NSE) measurement in hemolyzed serum samples
- PMID: 23778024
- DOI: 10.1016/j.cca.2013.06.009
Individualized correction of neuron-specific enolase (NSE) measurement in hemolyzed serum samples
Abstract
Background: Accuracy of serum neuron-specific enolase (NSE) measurement is paramount, particularly in the context of neurological outcome prognostication. However, NSE measurements are compromised by even slight hemolysis, as it is abundant in red blood cells (RBCs). We derived and validated an individualized hemolysis correction equation in an attempt to reduce the current rejection rate of 14% at our institution.
Methods: Intracellular NSE was measured in RBC lysates to determine concentration variability. A correction equation was derived, accounting for both RBC-derived NSE false-elevation and hemoglobin-derived signal quenching. The performance of this individualized correction was evaluated in intentionally hemolyzed samples and accuracy was compared to a generalized correction.
Results: Significant inter-individual variability of RBC NSE was observed, with an almost two-fold range (15.7-28.5 ng NSE/mg Hb, p<0.001); intra-individual variability was insignificant. The individualized hemolysis correction equation derived: NSE(corr)=NSE(meas)-(Hb(serum))(NSE(RBCs/Hb))+0.0844(Hb(serum))+1.1 corrected 95% of the intentionally hemolyzed samples to within ±5 ng/ml of corresponding baseline NSE concentrations, compared to 74% using a generalized formula.
Conclusions: The individualized hemolysis correction provides increased accuracy in the estimation of true serum NSE concentrations for hemolyzed samples, compared to a generalized approach, by accounting for inter-individual RBC NSE variability. Incorporating this correction should reduce sample rejection rates and overall health care costs.
Keywords: Individualized hemolysis correction; Ischemic brain injury; Neurological prognosis; Neuron-specific enolase; Therapeutic hypothermia.
Copyright © 2013 Elsevier B.V. All rights reserved.
Similar articles
-
Individualized correction of insulin measurement in hemolyzed serum samples.Immunol Res. 2017 Jun;65(3):605-608. doi: 10.1007/s12026-016-8878-7. Immunol Res. 2017. PMID: 27817149
-
The automated processing algorithm to correct the test result of serum neuron-specific enolase affected by specimen hemolysis.J Clin Lab Anal. 2021 Sep;35(9):e23895. doi: 10.1002/jcla.23895. Epub 2021 Jul 7. J Clin Lab Anal. 2021. PMID: 34233042 Free PMC article.
-
Derivation and validation of an equation for adjustment of neuron-specific enolase concentrations in hemolyzed serum.Pediatr Crit Care Med. 2009 Mar;10(2):260-3. doi: 10.1097/PCC.0b013e31819a376d. Pediatr Crit Care Med. 2009. PMID: 19188872
-
A correction formula for neuron-specific enolase measurement in hemolyzed neonatal serum samples.Biomed Rep. 2018 May;8(5):491-496. doi: 10.3892/br.2018.1081. Epub 2018 Mar 21. Biomed Rep. 2018. PMID: 29725527 Free PMC article.
-
Development and validation of a simple correction method for the measurement of neuron-specific enolase in hemolyzed serum samples.Scand J Clin Lab Invest. 2022 Feb;82(1):12-17. doi: 10.1080/00365513.2021.2009021. Epub 2021 Dec 8. Scand J Clin Lab Invest. 2022. PMID: 34878363
Cited by
-
Correction of the NSE concentration in hemolyzed serum samples improves its diagnostic accuracy in small-cell lung cancer.Oncotarget. 2020 Jul 7;11(27):2660-2668. doi: 10.18632/oncotarget.27664. eCollection 2020 Jul 7. Oncotarget. 2020. PMID: 32676167 Free PMC article.
-
Tasks and Experiences of the Prospective, Longitudinal, Multicenter MoMar (Molecular Markers) Study for the Early Detection of Mesothelioma in Individuals Formerly Exposed to Asbestos Using Liquid Biopsies.Cancers (Basel). 2023 Dec 18;15(24):5896. doi: 10.3390/cancers15245896. Cancers (Basel). 2023. PMID: 38136442 Free PMC article.
-
Circulating long non-coding RNA GAS5 (growth arrest-specific transcript 5) as a complement marker for the detection of malignant mesothelioma using liquid biopsies.Biomark Res. 2020 May 13;8:15. doi: 10.1186/s40364-020-00194-4. eCollection 2020. Biomark Res. 2020. PMID: 32435497 Free PMC article.
-
Individualized correction of insulin measurement in hemolyzed serum samples.Immunol Res. 2017 Jun;65(3):605-608. doi: 10.1007/s12026-016-8878-7. Immunol Res. 2017. PMID: 27817149
-
The automated processing algorithm to correct the test result of serum neuron-specific enolase affected by specimen hemolysis.J Clin Lab Anal. 2021 Sep;35(9):e23895. doi: 10.1002/jcla.23895. Epub 2021 Jul 7. J Clin Lab Anal. 2021. PMID: 34233042 Free PMC article.