Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2009 Dec;31(6):657-64.
doi: 10.1111/j.1751-553X.2008.01093.x. Epub 2008 Aug 28.

Point-of-care method for total white cell count: an evaluation of the HemoCue WBC device

Affiliations
Free PMC article

Point-of-care method for total white cell count: an evaluation of the HemoCue WBC device

A Osei-Bimpong et al. Int J Lab Hematol. 2009 Dec.
Free PMC article

Abstract

Point-of-care testing (POCT) is becoming an important adjunct to haematology laboratory practice. An important component of the blood count is the total white cell count (WBC). Previously, this required laborious microscopic cell counting, but it can now be performed by means of automation; however, in many under-resourced countries, costly automated counters are only available in very few central hospitals. Moreover, neither method is practical in most POCT situations. The HemoCue WBC has been developed as a simplified alternative method, consisting of a reagent pre-loaded disposable cuvette together with basic image analysis technology. This report describes an assessment of its utility. The WBC of 500 routine blood samples from the hospital were tested in parallel by the HemoCue WBC and by a reference analyser to assess accuracy and utility of the former. The tests included precision, linearity, type of blood sample and anticoagulant and potential interfering substances in blood specimens. In the tests for accuracy, 192 of the 200 showed percentage difference from the NEQAS reference of <10% whilst the remaining eight samples differed by <12%, thus meeting the requirements of Clinical laboratory improvement amendments (CLIA)-88 regulations. Of the samples tested with potential interfering substances only those with >2% normoblasts or reticulocytosis showed significant differences from the reference measurements. The HemoCue WBC is reliable for WBC counts within the analytical range of 0.4-30.0 x 10(9)/l, except in samples where there are significant numbers of normoblasts or reticulocytes. It is simple to use and provides a valuable advance in the facilities available for POCT in haematology.

PubMed Disclaimer

Figures

Figure 2
Figure 2
Microscopic appearance of a cuvette filled with a blood sample. (×40) Photographed on a conventional light microscope (Nikon E400; Nikon Electronic Company, Osaka, Japan).
Figure 1
Figure 1
Photograph of the portable HemoCue WBC Point-of-care analyser (plastic cuvette containing a drop of blood is shown placed on the circular base holder).
Figure 3
Figure 3
Correlation between the reference analyser and the HemoCue WBC. Y = 0.989X−0.082 (Y = reference analyser and X = HemoCue WBC). The correlation coefficient (r) between 0.4 × 109/l and 30 × 109/l = 0.997. These findings indicate good comparability within the manufacturer’s suggested analytic range with no detectable bias.
Figure 4
Figure 4
Comparison of HemoCue WBC measurements on capillary and venous blood samples:

Similar articles

Cited by

References

    1. Bates I, Mendelow B. Haematology in under-resourced laboratories. In: Lewis SM, Bain B, Bates I, editors. Dacie & Lewis Practical Haematology. Philadelphia, PA: Churchill Livingstone Elsevier; 2006. p. 674.
    1. Briggs C, Carter J, Lee S-H, Sandhaus L, Simon –Lopez R, Vives Corrons J-L. ICSH Guideline for worldwide Point of Care testing in haematology with special reference to the complete blood count. International Journal of Laboratory Haematology. 2008;30:105–116. - PubMed
    1. England JM, Hyde K, Lewis SM, Mackie IJ, Rowan RM. Guide-lines for near patient testing: haematology. Clinical and Laboratory Haematology. 1995;17:301–310. - PubMed
    1. International Council for Standardization in Haematology Guidelines for the evaluation of blood cell analysers including those used for differential leucocyte and reticulocyte counting and cell marker applications. Clinical and Laboratory Haematology. 1994;16:157–174. - PubMed
    1. Lewis SM, De la Salle B. Quality assurance. In: Lewis SM, Bain B, Bates I, editors. Dacie & Lewis Practical Haematology. Philadelphia, PA: Churchill Livingstone Elsevier; 2006. p. 669.

Publication types

LinkOut - more resources