Genotoxicity of corrosion eluates obtained from endosseous implants
- PMID: 17356376
- DOI: 10.1097/ID.0b013e3180327663
Genotoxicity of corrosion eluates obtained from endosseous implants
Abstract
Purpose: Commercially pure titanium alloys are currently used as metallic biomaterials in implantology. Corrosion phenomena appear to play a decisive role in metallic implant long-term behavior. Thus, the goal of this study was to examine the genotoxic potential of corrosion eluates obtained from dental implants using Chinese ovary hamster cells in vitro by the single-cell gel (comet) assay. This technique detects deoxyribonucleic acid strand breaks in individual cells in alkaline conditions.
Materials and methods: The materials tested included 3 dental implants commercially available. Each of the tested materials was corroded in a solution consisting of equal amounts of acetic acid and sodium chloride (0.1 M) for 1, 3, 7, 14, and 21 days. The Chinese ovary hamster cultures were then exposed to all corrosion eluates obtained from endosseous dental implants for 30 minutes at 37 degrees C.
Results: None of the eluates was found to exhibit genotoxicity, regardless of the type of dental implant used.
Conclusion: The results suggest that all dental implants tested in this study did not induce deoxyribonucleic acid breakage as depicted by the single-cell gel (comet) assay.
Similar articles
-
Genotoxicity of endosseous implants using two cellular lineages in vitro.J Oral Implantol. 2014 Feb;40(1):25-9. doi: 10.1563/AAID-JOI-D-10-00112. Epub 2011 Jun 8. J Oral Implantol. 2014. PMID: 21651414
-
Genotoxicity of corrosion eluates obtained from orthodontic brackets in vitro.Am J Orthod Dentofacial Orthop. 2011 Apr;139(4):504-9. doi: 10.1016/j.ajodo.2009.03.058. Am J Orthod Dentofacial Orthop. 2011. PMID: 21457861
-
Genotoxicity and cytotoxicity of mineral trioxide aggregate and regular and white Portland cements on Chinese hamster ovary (CHO) cells in vitro.Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2006 Feb;101(2):258-61. doi: 10.1016/j.tripleo.2005.02.080. Epub 2005 Oct 14. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2006. PMID: 16448930
-
The issue of corrosion in dental implants: a review.Acta Odontol Latinoam. 2009;22(1):3-9. Acta Odontol Latinoam. 2009. PMID: 19601489 Review.
-
Corrosion of implantable materials and aging of bone.J Dent Technol. 1996 Jun;13(5):24-32. J Dent Technol. 1996. PMID: 9516274 Review.
Cited by
-
Clinical Prospective Assessment of Genotoxic Effects of Dental Implants in Gingival Epithelial Cells.Acta Stomatol Croat. 2022 Sep;56(3):222-234. doi: 10.15644/asc56/3/1. Acta Stomatol Croat. 2022. PMID: 36382208 Free PMC article.
-
Tuning Surface Chemistry of Polyetheretherketone by Gold Coating and Plasma Treatment.Nanoscale Res Lett. 2017 Dec;12(1):424. doi: 10.1186/s11671-017-2182-x. Epub 2017 Jun 21. Nanoscale Res Lett. 2017. PMID: 28637351 Free PMC article.
-
The effect of platform switching on the levels of metal ion release from different implant-abutment couples.Int J Oral Sci. 2016 Jun 30;8(2):117-25. doi: 10.1038/ijos.2016.5. Int J Oral Sci. 2016. PMID: 27357323 Free PMC article.
-
Current strategies to improve the bioactivity of PEEK.Int J Mol Sci. 2014 Mar 28;15(4):5426-45. doi: 10.3390/ijms15045426. Int J Mol Sci. 2014. PMID: 24686515 Free PMC article. Review.
-
Metallic ion content and damage to the DNA in oral mucosa cells patients treated dental implants.J Mater Sci Mater Med. 2014 Jul;25(7):1819-24. doi: 10.1007/s10856-014-5203-7. Epub 2014 Mar 29. J Mater Sci Mater Med. 2014. PMID: 24682897