The long-term mechanical integrity of non-reinforced PEEK-OPTIMA polymer for demanding spinal applications: experimental and finite-element analysis
- PMID: 15940477
- PMCID: PMC3489413
- DOI: 10.1007/s00586-005-0915-5
The long-term mechanical integrity of non-reinforced PEEK-OPTIMA polymer for demanding spinal applications: experimental and finite-element analysis
Abstract
Polyetheretherketone (PEEK) is a novel polymer with potential advantages for its use in demanding orthopaedic applications (e.g. intervertebral cages). However, the influence of a physiological environment on the mechanical stability of PEEK has not been reported. Furthermore, the suitability of the polymer for use in highly stressed spinal implants such as intervertebral cages has not been investigated. Therefore, a combined experimental and analytical study was performed to address these open questions. A quasi-static mechanical compression test was performed to compare the initial mechanical properties of PEEK-OPTIMA polymer in a dry, room-temperature and in an aqueous, 37 degrees C environment (n=10 per group). The creep behaviour of cylindrical PEEK polymer specimens (n=6) was measured in a simulated physiological environment at an applied stress level of 10 MPa for a loading duration of 2000 hours (12 weeks). To compare the biomechanical performance of different intervertebral cage types made from PEEK and titanium under complex loading conditions, a three-dimensional finite element model of a functional spinal unit was created. The elastic modulus of PEEK polymer specimens in a physiological environment was 1.8% lower than that of specimens tested at dry, room temperature conditions (P<0.001). The results from the creep test showed an average creep strain of less than 0.1% after 2000 hours of loading. The finite element analysis demonstrated high strain and stress concentrations at the bone/implant interface, emphasizing the importance of cage geometry for load distribution. The stress and strain maxima in the implants were well below the material strength limits of PEEK. In summary, the experimental results verified the mechanical stability of the PEEK-OPTIMA polymer in a simulated physiological environment, and over extended loading periods. Finite element analysis supported the use of PEEK-OPTIMA for load-bearing intervertebral implants.
Figures
Similar articles
-
Biomechanical assessment of a PEEK rod system for semi-rigid fixation of lumbar fusion constructs.J Biomech Eng. 2011 Aug;133(8):081009. doi: 10.1115/1.4004862. J Biomech Eng. 2011. PMID: 21950902
-
Effect of porous orthopaedic implant material and structure on load sharing with simulated bone ingrowth: A finite element analysis comparing titanium and PEEK.J Mech Behav Biomed Mater. 2018 Apr;80:68-76. doi: 10.1016/j.jmbbm.2018.01.017. J Mech Behav Biomed Mater. 2018. PMID: 29414477 Free PMC article.
-
Biomechanical evaluation and comparison of polyetheretherketone rod system to traditional titanium rod fixation.Spine J. 2009 Mar;9(3):263-7. doi: 10.1016/j.spinee.2008.08.002. Epub 2008 Oct 1. Spine J. 2009. PMID: 18838341
-
PEEK dental implants: a review of the literature.J Oral Implantol. 2013 Dec;39(6):743-9. doi: 10.1563/AAID-JOI-D-11-00002. Epub 2011 Sep 9. J Oral Implantol. 2013. PMID: 21905892 Review.
-
Carbon fiber-reinforced PEEK in implant dentistry: A scoping review on the finite element method.Comput Methods Biomech Biomed Engin. 2021 Sep;24(12):1355-1367. doi: 10.1080/10255842.2021.1888939. Epub 2021 Feb 22. Comput Methods Biomech Biomed Engin. 2021. PMID: 33616450 Review.
Cited by
-
In silico medical device testing of anatomically and mechanically conforming patient-specific spinal fusion cages designed by full-scale topology optimisation.Front Bioeng Biotechnol. 2024 Sep 10;12:1347961. doi: 10.3389/fbioe.2024.1347961. eCollection 2024. Front Bioeng Biotechnol. 2024. PMID: 39318669 Free PMC article.
-
Low-Modulus PMMA Has the Potential to Reduce Stresses on Endplates after Cement Discoplasty.J Funct Biomater. 2022 Feb 4;13(1):18. doi: 10.3390/jfb13010018. J Funct Biomater. 2022. PMID: 35225981 Free PMC article.
-
Review on Polymer, Ceramic and Composite Materials for CAD/CAM Indirect Restorations in Dentistry-Application, Mechanical Characteristics and Comparison.Materials (Basel). 2021 Mar 24;14(7):1592. doi: 10.3390/ma14071592. Materials (Basel). 2021. PMID: 33805176 Free PMC article. Review.
-
Bending Behaviour of Polymeric Materials Used on Biomechanics Orthodontic Appliances.Materials (Basel). 2020 Dec 7;13(23):5579. doi: 10.3390/ma13235579. Materials (Basel). 2020. PMID: 33297572 Free PMC article.
-
Endplate Changes with Polyetheretherketone Cages in Posterior Lumbar Interbody Fusion.Asian Spine J. 2020 Apr;14(2):229-237. doi: 10.31616/asj.2019.0124. Epub 2019 Nov 12. Asian Spine J. 2020. PMID: 31711063 Free PMC article.
References
-
- Abu Bakar MS, Cheng MH, Tang SM, Yu SC, Liao K, Tan CT, Khor KA, Cheang P. Tensile properties, tension–tension fatigue and biological response of polyetheretherketone-hydroxyapatite composites for load-bearing orthopedic implants. Biomaterials. 2003;24:2245–2250. doi: 10.1016/S0142-9612(03)00028-0. - DOI - PubMed
-
- Akay M, Aslan N. An estimation of fatigue life for a carbon fibre/poly ether ether ketone hip joint prosthesis. Proc Inst Mech Eng [H] 1995;209:93–103. - PubMed
-
- Albert K, Schledjewski R, Harbaugh M, Bleser S, Jamison R, Friedrich K. Characterization of wear in composite material orthopaedic implants. Part II: The implant/bone interface. Biomed Mater Eng. 1994;4:199–211. - PubMed
-
- ASTM D2990–01 (2004) Standard test methods for tensile, compressive, and flexural creep and creep-rupture of plastics
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources