Biological response to wear debris generated in carbon based composites as potential bearing surfaces for artificial hip joints
- PMID: 14598403
- DOI: 10.1002/jbm.b.10068
Biological response to wear debris generated in carbon based composites as potential bearing surfaces for artificial hip joints
Abstract
UHMWPE wear particles have been implicated in osteolysis, implant loosening, and long-term failure of total hip arthroplasties in vivo. This study examined four carbon-based composite materials as alternatives for UHMWPE in joint bearings. These materials were HMU-CVD, SMS-CVD, P25-CVD, and CFR-PEEK. New bearing materials should satisfy certain criteria: they should have good wear properties that at least match UHMWPE, and produce wear particles with low levels of biological activity. Of the four materials tested in multidirectional pin-on-plate tribological tests, SMS-CVD, P25-CVD, and CFR-PEEK showed lower volumetric wear factors than UHMWPE. P25-CVD had the lowest wear factor of 0.54 +/- 0.34 x 10(-7) mm(3)/Nm. Analysis of P25-CVD wear particles by transmission electron microscopy showed that the debris was very small, with the vast majority of particles being under 100 nm in size, which was similar in size to metal wear particles. The P25-CVD particles were isolated and cultured with L929 fibroblasts and U937 monocytic cells to assess their effect on cell viability. P25-CVD particles were significantly less cytotoxic (p < 0.01, ANOVA) to both cell types than CoCr metal wear particles. This work suggests that carbon-carbon composite materials may have potential for use in total hip replacement bearings. Of the materials tested P25-CVD had the lowest wear factor, and produced very small wear debris that had minimal cytotoxic effect on L929 and U937 cells in vitro. Therefore carbon-carbon composites, such as P25-CVD, may be important in the development of next-generation implants with lower wear rates and reduced cytotoxic potential.
Copyright 2003 Wiley Periodicals, Inc.
Similar articles
-
Carbon-carbon composite bearing materials in hip arthroplasty: analysis of wear and biological response to wear debris.J Mater Sci Mater Med. 2004 Jan;15(1):91-8. doi: 10.1023/b:jmsm.0000010102.26218.d1. J Mater Sci Mater Med. 2004. PMID: 15338596
-
Comparison of the cytotoxicity of clinically relevant cobalt-chromium and alumina ceramic wear particles in vitro.Biomaterials. 2003 Feb;24(3):469-79. doi: 10.1016/s0142-9612(02)00360-5. Biomaterials. 2003. PMID: 12423602
-
The Biotribology of PEEK-on-HXLPE Bearings Is Comparable to Traditional Bearings on a Multidirectional Pin-on-disk Tester.Clin Orthop Relat Res. 2016 Nov;474(11):2384-2393. doi: 10.1007/s11999-016-4989-7. Clin Orthop Relat Res. 2016. PMID: 27457625 Free PMC article.
-
The Biologic Response to Polyetheretherketone (PEEK) Wear Particles in Total Joint Replacement: A Systematic Review.Clin Orthop Relat Res. 2016 Nov;474(11):2394-2404. doi: 10.1007/s11999-016-4976-z. Epub 2016 Jul 18. Clin Orthop Relat Res. 2016. PMID: 27432420 Free PMC article. Review.
-
Biological reactions to wear debris in total joint replacement.Proc Inst Mech Eng H. 2000;214(1):21-37. doi: 10.1243/0954411001535219. Proc Inst Mech Eng H. 2000. PMID: 10718048 Review.
Cited by
-
Histopathological Analysis of PEEK Wear Particle Effects on the Synovial Tissue of Patients.Biomed Res Int. 2016;2016:2198914. doi: 10.1155/2016/2198914. Epub 2016 Aug 16. Biomed Res Int. 2016. PMID: 27766256 Free PMC article.
-
Wear and Friction of UHMWPE-on-PEEK OPTIMA™.J Mech Behav Biomed Mater. 2019 Jan;89:65-71. doi: 10.1016/j.jmbbm.2018.09.021. Epub 2018 Sep 17. J Mech Behav Biomed Mater. 2019. PMID: 30265867 Free PMC article.
-
Long Term Results of a Rotating Hinge Total Knee Prosthesis With Carbon-Fiber Reinforced Poly-Ether-Ether-Ketone (CFR-PEEK) as Bearing Material.Front Bioeng Biotechnol. 2022 Mar 4;10:845859. doi: 10.3389/fbioe.2022.845859. eCollection 2022. Front Bioeng Biotechnol. 2022. PMID: 35356773 Free PMC article.
-
PEEK (Polyether-ether-ketone) Based Cervical Total Disc Arthroplasty: Contact Stress and Lubrication Analysis.Open Biomed Eng J. 2012;6:73-9. doi: 10.2174/1874230001206010073. Epub 2012 May 9. Open Biomed Eng J. 2012. PMID: 22670159 Free PMC article.
-
Significance of nano- and microtopography for cell-surface interactions in orthopaedic implants.J Biomed Biotechnol. 2007;2007(8):69036. doi: 10.1155/2007/69036. J Biomed Biotechnol. 2007. PMID: 18274618 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources