International investigation of breast MRI: results of a multicentre study (11 sites) concerning diagnostic parameters for contrast-enhanced MRI based on 519 histopathologically correlated lesions
- PMID: 11354744
- DOI: 10.1007/s003300000745
International investigation of breast MRI: results of a multicentre study (11 sites) concerning diagnostic parameters for contrast-enhanced MRI based on 519 histopathologically correlated lesions
Abstract
A multicentre study was undertaken to provide fundamentals for improved standardization and optimized interpretation guidelines of dynamic contrast-enhanced MRI. Only patients scheduled for biopsy of a clinical or imaging abnormality were included. They underwent standardized dynamic MRI on Siemens 1.0 (163 valid lesions > or = 5 mm) or 1.5 T (395 valid lesions > or = 5 mm) using 3D fast low-angle shot (FLASH; 87 s) before and five times after standardized bolus of 0.2 mmol Gd-DTPA/kg. One-Tesla and 1.5 T data were analysed separately using a discriminant analysis. Only histologically correlated lesions entered the statistical evaluation. Histopathology and imaging were correlated in retrospect and in open. The best results were achieved by combining up to five wash-in or wash-out parameters. Different weighting of false-negative vs false-positive calls allowed formulation of a statistically based interpretation scheme yielding optimized rules for the highest possible sensitivity (specificity 30%), for moderate (50%) or high (64-71%) specificity. The sensitivities obtained at the above specificity levels were better at 1.0 T (98, 97, or 96%) than at 1.5 T (96, 93, 86%). Using a widely available standardized MR technique definition of statistically founded interpretation rules is possible. Choice of an optimum interpretation rule may vary with the clinical question. Prospective testing remains necessary. Differences of 1.0 and 1.5 T are not statistically significant but may be due to pulse sequences.
Similar articles
-
Benign and malignant breast lesions: efficacy of real time contrast-enhanced ultrasound vs. magnetic resonance imaging.Ultraschall Med. 2007 Feb;28(1):57-62. doi: 10.1055/s-2006-927226. Ultraschall Med. 2007. PMID: 17304413
-
Comparison of gadoteric acid and gadobutrol for detection as well as morphologic and dynamic characterization of lesions on breast dynamic contrast-enhanced magnetic resonance imaging.Invest Radiol. 2014 Jul;49(7):474-84. doi: 10.1097/RLI.0000000000000039. Invest Radiol. 2014. PMID: 24637587 Clinical Trial.
-
The current status of breast MR imaging. Part I. Choice of technique, image interpretation, diagnostic accuracy, and transfer to clinical practice.Radiology. 2007 Aug;244(2):356-78. doi: 10.1148/radiol.2442051620. Radiology. 2007. PMID: 17641361 Review.
-
[Combination of low and high resolution T1-weighted sequences for improved evaluation of morphologic criteria in dynamic contrast enhanced MRI of the breast].Rofo. 2002 Nov;174(11):1445-9. doi: 10.1055/s-2002-35350. Rofo. 2002. PMID: 12424673 German.
-
Contrast-enhanced MRI of the breast: accuracy, value, controversies, solutions.Eur J Radiol. 1997 Feb;24(2):94-108. doi: 10.1016/s0720-048x(96)01142-4. Eur J Radiol. 1997. PMID: 9097051 Review.
Cited by
-
Kinetic curves of malignant lesions are not consistent across MRI systems: need for improved standardization of breast dynamic contrast-enhanced MRI acquisition.AJR Am J Roentgenol. 2009 Sep;193(3):832-9. doi: 10.2214/AJR.08.2025. AJR Am J Roentgenol. 2009. PMID: 19696299 Free PMC article.
-
Size assessment of breast lesions by means of a computer-aided detection (CAD) system for magnetic resonance mammography.Radiol Med. 2011 Oct;116(7):1039-49. doi: 10.1007/s11547-011-0664-y. Epub 2011 Mar 19. Radiol Med. 2011. PMID: 21424564 English, Italian.
-
Breast cancer preoperative staging: does contrast-enhanced magnetic resonance mammography modify surgery?Int J Breast Cancer. 2011;2011:757234. doi: 10.4061/2011/757234. Epub 2011 Jul 3. Int J Breast Cancer. 2011. PMID: 22295233 Free PMC article.
-
Comparison of postoperative CT- and preoperative MRI-based breast tumor bed contours in prone position for radiotherapy after breast-conserving surgery.Eur Radiol. 2021 Jan;31(1):345-355. doi: 10.1007/s00330-020-07085-0. Epub 2020 Aug 1. Eur Radiol. 2021. PMID: 32740818 Free PMC article.
-
Development and characterization of a dynamic lesion phantom for the quantitative evaluation of dynamic contrast-enhanced MRI.Med Phys. 2011 Oct;38(10):5601-11. doi: 10.1118/1.3633911. Med Phys. 2011. PMID: 21992378 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical