Performance of virus isolation and Directigen Flu A to detect influenza A virus in experimental human infection
- PMID: 10614856
- DOI: 10.1016/s1386-6532(99)00058-x
Performance of virus isolation and Directigen Flu A to detect influenza A virus in experimental human infection
Abstract
Background: Few data exist to assess the sensitivity of different specimen types for viral detection during the course of influenza virus infection.
Objectives: This study assessed the relationships between quantitative influenza A virus replication and antigen detectability by the enzyme immunosorbent assay (EIA) Directigen Flu A in different type of samples during experimental human infection.
Study design: Fourteen volunteers were inoculated with influenza A virus A/Texas/36/91 (H1N1). Four specimens types were collected in sequence for quantitative isolation in cell culture and antigen testing from days 1 to 8 after inoculation.
Results: Seventy-one (63%) of nasopharyngeal wash specimens were culture positive, compared to 51 (46%) of throat gargles, 51 (46%) of nasal swabs, and 27 (24%) of throat swabs. All subjects shed virus in their nasopharyngeal wash at least one day and 86% of subjects had a positive nasopharyngeal wash culture on day 2 after inoculation. The mean viral titers were highest on day 2 post inoculation for all specimen types and averaged 3.6 log10 TCID50/ml for nasal washes, 1.2 log10 TCID50/ml for throat gargles, 1.8 log10 TCID50/ml for the nasopharyngeal swabs, and 0.6 log10 TCID50/ml for the throat swabs. Mean viral titers in the nasal washes were significantly different (P<0.05) compared to other specimen types. The peak of sensitivity of EIA (compared to culture) was the second day after inoculation. Nasopharyngeal and throat swab results were combined for this analysis and considered positive by culture if positive in either or both samples. Thus, on day 2 the number of EIA positive samples relative to the number culture positive was 9/12 (75%) for nasopharyngeal wash specimens, 2/9 (22%) for throat gargles, and 7/11 (64%) for the combined throat and nasal swabs specimens.
Conclusions: Nasopharyngeal washes are the most sensitive sample type detecting influenza A virus in adults. For rapid diagnosis the Directigen Flu A is an alternative with a sensitivity compared to culture ranging between 64 and 78% if performed on nasopharyngeal specimens on day two or three after experimental infection in adults. However, if performed on other specimens or later in the course of infection the sensitivity is lower.
Similar articles
-
Application of Directigen FLU-A for the detection of influenza A virus in human and nonhuman specimens.J Clin Microbiol. 1992 May;30(5):1072-5. doi: 10.1128/jcm.30.5.1072-1075.1992. J Clin Microbiol. 1992. PMID: 1583103 Free PMC article.
-
Rapid diagnosis of influenza type A infection: comparison of shell-vial culture, directigen flu-A and enzyme-linked immunosorbent assay.New Microbiol. 1996 Apr;19(2):141-7. New Microbiol. 1996. PMID: 8722310
-
[Evaluation of a rapid enzyme immunoassay membrane test for diagnosis of influenza A virus infection].Kansenshogaku Zasshi. 1998 Aug;72(8):827-33. doi: 10.11150/kansenshogakuzasshi1970.72.827. Kansenshogaku Zasshi. 1998. PMID: 9780586 Japanese.
-
Laboratory diagnosis of influenza: recent advances.Semin Pediatr Infect Dis. 2002 Apr;13(2):85-9. doi: 10.1053/spid.2002.122994. Semin Pediatr Infect Dis. 2002. PMID: 12122957 Review.
-
Nasopharyngeal Swabs vs. Nasal Aspirates for Respiratory Virus Detection: A Systematic Review.Pathogens. 2021 Nov 20;10(11):1515. doi: 10.3390/pathogens10111515. Pathogens. 2021. PMID: 34832670 Free PMC article. Review.
Cited by
-
Clinical evaluation of the ZstatFlu-II test: a chemiluminescent rapid diagnostic test for influenza virus.J Clin Microbiol. 2002 Jul;40(7):2331-4. doi: 10.1128/JCM.40.7.2331-2334.2002. J Clin Microbiol. 2002. PMID: 12089243 Free PMC article.
-
Influenza.Lancet. 2003 Nov 22;362(9397):1733-45. doi: 10.1016/S0140-6736(03)14854-4. Lancet. 2003. PMID: 14643124 Free PMC article. Review.
-
A microfluidic immunomagnetic bead-based system for the rapid detection of influenza infections: from purified virus particles to clinical specimens.Biomed Microdevices. 2013 Jun;15(3):539-51. doi: 10.1007/s10544-013-9753-0. Biomed Microdevices. 2013. PMID: 23420191 Free PMC article.
-
Rapid Tests for Influenza.Curr Infect Dis Rep. 2005 May;7(3):187-192. doi: 10.1007/s11908-005-0033-x. Curr Infect Dis Rep. 2005. PMID: 15847720
-
Detection of SARS coronavirus in patients with suspected SARS.Emerg Infect Dis. 2004 Feb;10(2):294-9. doi: 10.3201/eid1002.030610. Emerg Infect Dis. 2004. PMID: 15030700 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical