Optimizing contrast enhancement during helical CT of the liver: a comparison of two bolus tracking techniques
- PMID: 9843287
- DOI: 10.2214/ajr.171.6.9843287
Optimizing contrast enhancement during helical CT of the liver: a comparison of two bolus tracking techniques
Abstract
Objective: The purpose of this study was to evaluate a recently developed hardware and software system for CT scanning that generates images in real time and switches to helical CT scanning by either a visual cue or a region of interest (ROI) amplitude threshold.
Subjects and methods: We randomly and prospectively divided 120 abdominal CT examinations into three groups. Two groups received 75 ml of contrast agent at 1.5 ml/sec. Helical CT scanning began after visualization of the contrast bolus arrival in the hepatic veins (visual cue triggering) (39 patients) or after reaching an ROI threshold (automated ROI threshold triggering) (39 patients). A third group served as a control group (42 patients) and received 150 ml of contrast agent at 1 ml/sec. Quality of hepatic enhancement was assessed objectively and subjectively. Comparisons were made after stratifying each group into three weight classes.
Results: Errors occurred in 12 (31%) of 39 examinations in the group with automated ROI threshold triggering. In that group, we found a significantly (p < .04) lower mean hepatic enhancement in two of three weight categories, and a significantly (p < .04) lower mean subjective scan quality in one of three weight categories, than we found in the group with visual cue triggering.
Conclusion: Optimizing portal venous phase helical CT of the liver after a low-volume bolus of contrast agent and an injection rate of 1.5 ml/sec is best achieved by initiating helical CT scanning after visualizing the contrast bolus arrival within the liver rather than after reaching a preset attenuation threshold.
Similar articles
-
[Effect of different threshold levels and injection rates on hepatic enhancement by using bolus tracking technique in helical CT].Tani Girisim Radyol. 2003 Jun;9(2):207-12. Tani Girisim Radyol. 2003. PMID: 14661491 Clinical Trial. Turkish.
-
Use of contrast material for spiral CT of the abdomen: comparison of hepatic enhancement and vascular attenuation for three different contrast media at two different delay times.AJR Am J Roentgenol. 1995 Feb;164(2):327-31. doi: 10.2214/ajr.164.2.7839963. AJR Am J Roentgenol. 1995. PMID: 7839963 Clinical Trial.
-
Multiple-phase helical CT of the liver for detecting small hepatomas in patients with liver cirrhosis: contrast-injection protocol and optimal timing.AJR Am J Roentgenol. 1996 Sep;167(3):753-7. doi: 10.2214/ajr.167.3.8751695. AJR Am J Roentgenol. 1996. PMID: 8751695 Clinical Trial.
-
CT scan of the liver.Radiol Clin North Am. 1998 Mar;36(2):247-61. doi: 10.1016/s0033-8389(05)70020-8. Radiol Clin North Am. 1998. PMID: 9520980 Review.
-
[Helical CT of the liver. Principles and applications].Gastroenterol Clin Biol. 1998 Jun-Jul;22(6-7):576-89. Gastroenterol Clin Biol. 1998. PMID: 9762328 Review. French. No abstract available.
Cited by
-
Contrast-to-Noise Ratio Optimization in Coronary Computed Tomography Angiography: Validation in a Swine Model.Acad Radiol. 2019 Jun;26(6):e115-e125. doi: 10.1016/j.acra.2018.06.026. Epub 2018 Aug 30. Acad Radiol. 2019. PMID: 30172714 Free PMC article.
-
Triple-phase computed tomography during arterial portography with bolus tracking for hepatic tumors.Jpn J Radiol. 2010 Feb;28(2):149-56. doi: 10.1007/s11604-009-0399-z. Epub 2010 Feb 26. Jpn J Radiol. 2010. PMID: 20182850
-
Study of an adaptive bolus chasing CT angiography.J Xray Sci Technol. 2006;14(1):27-38. J Xray Sci Technol. 2006. PMID: 27034539 Free PMC article.
-
Patency of cavopulmonary connection studied by single phase electron beam computed tomography.Int J Cardiovasc Imaging. 2003 Oct;19(5):447-55. doi: 10.1023/a:1025808900239. Int J Cardiovasc Imaging. 2003. PMID: 14609197
-
Adaptive Bolus Chasing Computed Tomography Angiography: Control Scheme and Experimental Results.Biomed Signal Process Control. 2008 Oct;3(4):319-326. doi: 10.1016/j.bspc.2008.04.005. Biomed Signal Process Control. 2008. PMID: 19802329 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical