Single unit recording capabilities of a 100 microelectrode array
- PMID: 8836553
Single unit recording capabilities of a 100 microelectrode array
Abstract
We have developed a three-dimensional silicon electrode array which provides 100 separate channels for neural recording in cortex. The device is manufactured using silicon micromachining techniques, and we have conducted acute recording experiments in cat striate cortex to evaluate the recording capabilities of the array. In a series of five acute experiments, 58.6% of the electrodes in the array were found to be capable of recording visually evoked responses. In the most recent acute study, the average signal-to-noise ratio for recordings obtained from 56 of the electrodes in the array was calculated to be 5.5:1. Using standard window discrimination techniques, an average of 3.4 separable spikes were identified for each of these electrodes. In order to compare the two-dimensional mapping capabilities of the array with those derived from other technologies, orientation preference and ocular dominance maps were generated for each of the evoked responses. Histological evaluation of the implant site indicates some localized tissue insult, but this is likely due to the perfusion procedure since high signal-to-noise ratio neural responses were recorded. The recording capabilities of the Utah Intracortical Electrode Array in combination with the large number of electrodes available for recording make the array a tool well suited for investigations into the parallel processing mechanisms in cortex.
Similar articles
-
Voltage-sensitive-dye imaging of microstimulation-evoked neural activity through intracortical horizontal and callosal connections in cat visual cortex.J Neural Eng. 2009 Dec;6(6):066002. doi: 10.1088/1741-2560/6/6/066002. Epub 2009 Oct 1. J Neural Eng. 2009. PMID: 19794238
-
Polytrodes: high-density silicon electrode arrays for large-scale multiunit recording.J Neurophysiol. 2005 May;93(5):2987-3000. doi: 10.1152/jn.01023.2004. Epub 2004 Nov 17. J Neurophysiol. 2005. PMID: 15548620
-
Reliability of signals from a chronically implanted, silicon-based electrode array in non-human primate primary motor cortex.IEEE Trans Neural Syst Rehabil Eng. 2005 Dec;13(4):524-41. doi: 10.1109/TNSRE.2005.857687. IEEE Trans Neural Syst Rehabil Eng. 2005. PMID: 16425835
-
Chronic multiunit recordings in behaving animals: advantages and limitations.Prog Brain Res. 2005;147:263-82. doi: 10.1016/S0079-6123(04)47020-4. Prog Brain Res. 2005. PMID: 15581712 Review.
-
Studies of human visual pathophysiology with visual evoked potentials.Clin Neurophysiol. 2006 Jul;117(7):1414-33. doi: 10.1016/j.clinph.2006.01.004. Epub 2006 Mar 3. Clin Neurophysiol. 2006. PMID: 16516551 Review.
Cited by
-
Design, Fabrication, and Implantation of Invasive Microelectrode Arrays as in vivo Brain Machine Interfaces: A Comprehensive Review.J Manuf Process. 2024 Sep 30;126:185-207. doi: 10.1016/j.jmapro.2024.07.100. Epub 2024 Jul 31. J Manuf Process. 2024. PMID: 39185373
-
The speech neuroprosthesis.Nat Rev Neurosci. 2024 Jul;25(7):473-492. doi: 10.1038/s41583-024-00819-9. Epub 2024 May 14. Nat Rev Neurosci. 2024. PMID: 38745103 Review.
-
Multiformity of extracellular microelectrode recordings from Aδ neurons in the dorsal root ganglia: a computational modeling study.J Neurophysiol. 2024 Feb 1;131(2):261-277. doi: 10.1152/jn.00385.2023. Epub 2024 Jan 3. J Neurophysiol. 2024. PMID: 38169334
-
Excitation creates a distributed pattern of cortical suppression due to varied recurrent input.Neuron. 2023 Dec 20;111(24):4086-4101.e5. doi: 10.1016/j.neuron.2023.09.010. Epub 2023 Oct 20. Neuron. 2023. PMID: 37865083
-
Manufacturing Processes of Implantable Microelectrode Array for In Vivo Neural Electrophysiological Recordings and Stimulation: A State-Of-the-Art Review.J Micro Nanomanuf. 2022 Dec 1;10(4):041001. doi: 10.1115/1.4063179. Epub 2023 Oct 9. J Micro Nanomanuf. 2022. PMID: 37860671 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous