Laboratory-based evaluation of the 4th-generation AlereTM HIV Combo rapid point-of-care test
- PMID: 38394120
- PMCID: PMC10889622
- DOI: 10.1371/journal.pone.0298912
Laboratory-based evaluation of the 4th-generation AlereTM HIV Combo rapid point-of-care test
Abstract
Background: Mozambique is a high-prevalence country for HIV and early detection of new HIV infections is crucial for control of the epidemic. We aimed to evaluate the accuracy of the 4th-generation rapid diagnostic test (RDT) AlereTM HIV Combo in detecting acute and seroconverted HIV-infection, among sexually-active women attending three clinical health centers in Maputo, Mozambique.
Methods: Women aged 14-55 years (n = 920) seeking care at the Mavalane Health Area, Maputo (February 2018-January 2019) were included, and blood specimens sampled. Sociodemographic and sexual behavior data were collected. Point-of-care HIV testing was performed using Alere DetermineTM HIV-1/2 and Uni-GoldTM HIV-1/2. All samples were also tested using Enzygnost® HIV Integral 4 and Innotest® HIV Antigen mAb in laboratory. The 4th-generation RDT AlereTM HIV Combo was evaluated on serum samples in the laboratory. Finally, Innotest® HIV Antigen mAb, Enzygnost® HIV Integral 4 (Ag/Ab), and HIV RNA quantification acted as gold standard assays in the evaluation of AlereTM HIV Combo test for HIV antigen detection (in clinical samples and in three HIV-1 seroconversion panels).
Results: The antibody component of the 4th generation AlereTM HIV Combo RDT demonstrated a sensitivity and specificity of 100% examining clinical samples. However, the test did not detect HIV p24 antigen in any clinical samples, while Innotest® HIV Antigen mAb, verified by Enzygnost® HIV Integral 4 (Ag/Ab) and/or HIV RNA quantification, detected HIV antigen in six clinical samples. Furthermore, the AlereTM HIV Combo RDT had a low sensitivity in the detection of HIV p24 antigen in seroconversion panels. The HIV prevalence among the examined women was 17.8%.
Conclusions: The 4th-generation RDT AlereTM HIV Combo showed similar sensitivity to the 3rd-generation RDTs to detect seroconverted HIV-infections. However, the sensitivity for detection of HIV p24 antigen and diagnosing acute HIV infections, before seroconversion, was low. There is an urgent need to develop and evaluate simple and affordable POC tests with high sensitivity and specificity for diagnosing individuals with acute HIV infection in resource-limited settings with high HIV prevalence.
Copyright: © 2024 Manjate et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Conflict of interest statement
The authors have declared that no competing interests exist.
Similar articles
-
Performance of the Alere Determine™ HIV-1/2 Ag/Ab Combo Rapid Test with specimens from HIV-1 seroconverters from the US and HIV-2 infected individuals from Ivory Coast.J Clin Virol. 2013 Dec;58 Suppl 1(Suppl 1):e54-8. doi: 10.1016/j.jcv.2013.07.002. Epub 2013 Aug 2. J Clin Virol. 2013. PMID: 23911678 Free PMC article.
-
Performance of serological and molecular tests within acute HIV infection.J Clin Virol. 2017 Aug;93:81-84. doi: 10.1016/j.jcv.2017.03.010. Epub 2017 Mar 21. J Clin Virol. 2017. PMID: 28351688
-
Performance comparison of the 4th generation Bio-Rad Laboratories GS HIV Combo Ag/Ab EIA on the EVOLIS™ automated system versus Abbott ARCHITECT HIV Ag/Ab Combo, Ortho Anti-HIV 1+2 EIA on Vitros ECi and Siemens HIV-1/O/2 enhanced on Advia Centaur.J Clin Virol. 2013 Dec;58 Suppl 1:e79-84. doi: 10.1016/j.jcv.2013.08.009. J Clin Virol. 2013. PMID: 24342482
-
Point-of-care tests detecting HIV nucleic acids for diagnosis of HIV-1 or HIV-2 infection in infants and children aged 18 months or less.Cochrane Database Syst Rev. 2021 Aug 12;8(8):CD013207. doi: 10.1002/14651858.CD013207.pub2. Cochrane Database Syst Rev. 2021. PMID: 34383961 Free PMC article. Review.
-
[Screening tests combined with p24 antigen and anti-HIV antibodies in early detection of HIV-1].Transfus Clin Biol. 2000 Jun;7 Suppl 1:18s-24s. doi: 10.1016/s1246-7820(00)80011-7. Transfus Clin Biol. 2000. PMID: 10919219 Review. French.
References
-
- UNAIDS, In Danger. Global AIDS Updated, 2022. Geneva, Switzerland. In: [Internet]. Publisher; 27 Jul 2022 [cited 11 Sept 2023]. Available from: https://www.unaids.org/en/resources/documents/2022/in-danger-global-aids....
-
- Parker E, Judge MA, Macete E, Nhampossa T, Dorward J, Langa DC, et al.. HIV infection in Eastern and Southern Africa: Highest burden, largest challenges, greatest potential. South Afr J HIV Med. 2021; 22(1):1237. Epub 2021/05/28. doi: 10.4102/sajhivmed.v22i1.1237 ; PubMed Central PMCID: PMC8182467. - DOI - PMC - PubMed
-
- UNDAIS. Fact sheet—Latest global and regional statistics on the status of the AIDS epidemic 2021. In: [Internet]. Publisher; 29 Nov 2021 [cited 11 Sept 2023]. Available from: https://www.unaids.org/en/resources/fact-sheet.
-
- UNAIDS, 2021 Evaluation of the UN Joint programme on HIV Mozambique 2016–2020. In: [Internet]. Publisher; 27 Aug 2021 [cited 11 Sept 2023]. Available from: https://www.unaids.org/sites/default/files/media_asset/evaluation-UN-Joi....
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Medical