A Review on the Biomechanical Effects of Fatigue on the Porcine Bioprosthetic Heart Valve
- PMID: 29773039
- DOI: 10.1615/JLongTermEffMedImplants.v27.i2-4.60
A Review on the Biomechanical Effects of Fatigue on the Porcine Bioprosthetic Heart Valve
Abstract
Characterization of the mechanisms of degeneration of porcine bioprosthetic heart valves (BHV) during long-term cyclic loading is required for predicting and ultimately preventing their failure. Isolation of purely mechanical effects from host biological ones is a necessary first step in understanding the fatigue process as a whole. Thus, in this review we focus on mechanical factors alone as a means of isolating their role in altering biomechanical properties and ultimately their contribution to the fatigue damage process. Mechanical evaluations included tension controlled biaxial, 3-point flexural, and uniaxial failure tests performed on cuspal tissue following 0, 50, 100, 200, and 300 × 106 in vitro accelerated test cycles. Overall, biaxial mechanical results indicate a decreasing radial extensibility that can be explained by stiffening of the effective collagen fiber network as well as a small decrease in the splay of the collagen fibers. Moreover, these results suggest that the loss in flexural rigidity with fatigue that we have previously measured (ASAIO 1999; 45:59-63) may not be because of loss of collagen stiffness alone, but also to fiber debonding and degradation of the amorphous extracellular matrix. We discuss the implications of these results that point toward the development of chemical-treatment methods that seek to maintain the integrity of the amorphous extracellular matrix to ultimately extend BHV long-term durability.
Similar articles
-
The biomechanical effects of fatigue on the porcine bioprosthetic heart valve.J Long Term Eff Med Implants. 2001;11(3-4):231-47. J Long Term Eff Med Implants. 2001. PMID: 11921666 Review.
-
Effects of fixation pressure on the biaxial mechanical behavior of porcine bioprosthetic heart valves with long-term cyclic loading.Biomaterials. 2002 Jun;23(11):2389-99. doi: 10.1016/s0142-9612(01)00375-1. Biomaterials. 2002. PMID: 12013187
-
Effects of cyclic flexural fatigue on porcine bioprosthetic heart valve heterograft biomaterials.J Biomed Mater Res A. 2010 Jul;94(1):205-13. doi: 10.1002/jbm.a.32659. J Biomed Mater Res A. 2010. PMID: 20166221 Free PMC article.
-
Effects of mechanical fatigue on the bending properties of the porcine bioprosthetic heart valve.ASAIO J. 1999 Jan-Feb;45(1):59-63. doi: 10.1097/00002480-199901000-00014. ASAIO J. 1999. PMID: 9952009
-
High-resolution magnetic resonance imaging to characterize the geometry of fatigued porcine bioprosthetic heart valves.J Heart Valve Dis. 1997 Jul;6(4):424-32. J Heart Valve Dis. 1997. PMID: 9263876 Review.
Cited by
-
Degeneration of Bioprosthetic Heart Valves: Update 2020.J Am Heart Assoc. 2020 Oct 20;9(19):e018506. doi: 10.1161/JAHA.120.018506. Epub 2020 Sep 21. J Am Heart Assoc. 2020. PMID: 32954917 Free PMC article. Review.
-
Proteolytic Degradation Is a Major Contributor to Bioprosthetic Heart Valve Failure.J Am Heart Assoc. 2023 Jan 3;12(1):e028215. doi: 10.1161/JAHA.122.028215. Epub 2022 Dec 24. J Am Heart Assoc. 2023. PMID: 36565196 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources