Review
. 2018 Apr;41(2):375-389.
doi: 10.1007/s10143-016-0763-2. Epub 2016 Jul 8.

Clinical and biomechanical researches of polyetheretherketone (PEEK) rods for semi-rigid lumbar fusion: a systematic review

Affiliations
Review

Clinical and biomechanical researches of polyetheretherketone (PEEK) rods for semi-rigid lumbar fusion: a systematic review

Chan Li et al. Neurosurg Rev. 2018 Apr.

Abstract

Lumbar spinal fusion using rigid rods is a common surgical technique. However, adjacent segment disease and other adverse effects can occur. Dynamic stabilization devices preserve physiologic motion and reduce painful stress but have a high rate of construct failure and reoperation. Polyetheretherketone (PEEK) rods for semi-rigid fusions have a similar stiffness and adequate stabilization power compared with titanium rods, but with improved load sharing and reduced mechanical failure. The purpose of this paper is to review and evaluate the clinical and biomechanical performance of PEEK rods. A systematic review of clinical and biomechanical studies was conducted. A literature search using the PubMed, EMBASE, and Cochrane Library databases identified studies that met the eligibility criteria. Eight clinical studies and 15 biomechanical studies were included in this systematic review. The visual analog scale and the Oswestry disability index improved significantly in most studies, with satisfactory fusion rates. The occurrence of adjacent segment disease was low. In biomechanical studies, PEEK rods demonstrated a superior load-sharing distribution, a larger adjacent segment range of motion, and reduced stress at the rod-screw/screw-bone interfaces compared with titanium rods. The PEEK rod construct was simple to assemble and had a reliable in vivo performance compared with dynamic devices. The quality of clinical studies was low with confounding results, although results from mechanical studies were encouraging. There is no evidence strong enough to confirm better outcomes with PEEK rods than titanium rods. More studies with better protocols, a larger sample size, and a longer follow-up time are needed.

Keywords: Adjacent segment disease; Fusion rates; Polyetheretherketone (PEEK) rods; Semi-rigid fusion.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Spine J. 2013 Jan;13(1):85-94 - PubMed
    1. Spine (Phila Pa 1976). 2012 Oct 15;37(22 Suppl):S133-43 - PubMed
    1. BMJ. 2008 Apr 26;336(7650):924-6 - PubMed
    1. Adv Orthop. 2013;2013:745610 - PubMed
    1. Biomed Mater. 2008 Dec;3(4):044101 - PubMed

LinkOut - more resources