Clinical and biomechanical researches of polyetheretherketone (PEEK) rods for semi-rigid lumbar fusion: a systematic review
- PMID: 27392677
- DOI: 10.1007/s10143-016-0763-2
Clinical and biomechanical researches of polyetheretherketone (PEEK) rods for semi-rigid lumbar fusion: a systematic review
Abstract
Lumbar spinal fusion using rigid rods is a common surgical technique. However, adjacent segment disease and other adverse effects can occur. Dynamic stabilization devices preserve physiologic motion and reduce painful stress but have a high rate of construct failure and reoperation. Polyetheretherketone (PEEK) rods for semi-rigid fusions have a similar stiffness and adequate stabilization power compared with titanium rods, but with improved load sharing and reduced mechanical failure. The purpose of this paper is to review and evaluate the clinical and biomechanical performance of PEEK rods. A systematic review of clinical and biomechanical studies was conducted. A literature search using the PubMed, EMBASE, and Cochrane Library databases identified studies that met the eligibility criteria. Eight clinical studies and 15 biomechanical studies were included in this systematic review. The visual analog scale and the Oswestry disability index improved significantly in most studies, with satisfactory fusion rates. The occurrence of adjacent segment disease was low. In biomechanical studies, PEEK rods demonstrated a superior load-sharing distribution, a larger adjacent segment range of motion, and reduced stress at the rod-screw/screw-bone interfaces compared with titanium rods. The PEEK rod construct was simple to assemble and had a reliable in vivo performance compared with dynamic devices. The quality of clinical studies was low with confounding results, although results from mechanical studies were encouraging. There is no evidence strong enough to confirm better outcomes with PEEK rods than titanium rods. More studies with better protocols, a larger sample size, and a longer follow-up time are needed.
Keywords: Adjacent segment disease; Fusion rates; Polyetheretherketone (PEEK) rods; Semi-rigid fusion.
Similar articles
-
Biomechanical assessment of a PEEK rod system for semi-rigid fixation of lumbar fusion constructs.J Biomech Eng. 2011 Aug;133(8):081009. doi: 10.1115/1.4004862. J Biomech Eng. 2011. PMID: 21950902
-
Polyetheretherketone (PEEK) Rods in Lumbar Spine Degenerative Disease: A Case Series.Clin Spine Surg. 2016 Aug;29(7):E371-5. doi: 10.1097/BSD.0b013e318277cb9b. Clin Spine Surg. 2016. PMID: 23075859
-
Biomechanical evaluation and comparison of polyetheretherketone rod system to traditional titanium rod fixation.Spine J. 2009 Mar;9(3):263-7. doi: 10.1016/j.spinee.2008.08.002. Epub 2008 Oct 1. Spine J. 2009. PMID: 18838341
-
Biomechanical and clinical research of Isobar semi-rigid stabilization devices for lumbar degenerative diseases: a systematic review.Biomed Eng Online. 2023 Sep 23;22(1):95. doi: 10.1186/s12938-023-01156-1. Biomed Eng Online. 2023. PMID: 37742006 Free PMC article. Review.
-
PEEK rod systems for the spine.Eur J Orthop Surg Traumatol. 2014 Jul;24 Suppl 1:S111-6. doi: 10.1007/s00590-014-1421-4. Epub 2014 Feb 2. Eur J Orthop Surg Traumatol. 2014. PMID: 24487665 Review.
Cited by
-
Satisfactory 2-year outcome of minimal invasive hybrid stabilization with double treated screws for unstable osteoporotic spinal fractures.Eur J Trauma Emerg Surg. 2024 Aug 30. doi: 10.1007/s00068-024-02645-1. Online ahead of print. Eur J Trauma Emerg Surg. 2024. PMID: 39212727
-
Comparison of biomechanical effects of polyetheretherketone (PEEK) rods and titanium rods in lumbar long-segment instrumentation: a finite element study.Front Bioeng Biotechnol. 2024 Jul 11;12:1416046. doi: 10.3389/fbioe.2024.1416046. eCollection 2024. Front Bioeng Biotechnol. 2024. PMID: 39055340 Free PMC article.
-
Biomechanical comparison of polyetheretherketone rods and titanium alloy rods in transforaminal lumbar interbody fusion: a finite element analysis.BMC Surg. 2024 May 29;24(1):169. doi: 10.1186/s12893-024-02462-8. BMC Surg. 2024. PMID: 38811965 Free PMC article.
-
Innovative technologies in thoracolumbar and lumbar spine surgery failing to reach standard of care: state-of-art review.Spine Deform. 2024 May 25. doi: 10.1007/s43390-024-00898-9. Online ahead of print. Spine Deform. 2024. PMID: 38795313 Review.
-
Regional differences in bone mineral density biomechanically induce a higher risk of adjacent vertebral fracture after percutaneous vertebroplasty: a case-comparative study.Int J Surg. 2023 Mar 1;109(3):352-363. doi: 10.1097/JS9.0000000000000273. Int J Surg. 2023. PMID: 36912508 Free PMC article.
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous