Application of quantitative immunofluorescence to clinical serology: antibody levels of Toxoplasma gondii
- PMID: 2674197
- PMCID: PMC267728
- DOI: 10.1128/jcm.27.9.2008-2013.1989
Application of quantitative immunofluorescence to clinical serology: antibody levels of Toxoplasma gondii
Abstract
Methods currently used for immunofluorescent reagent standardization require subjective visual comparison of reagents with control materials. Reactivities of reagents in immunofluorescence test kits vary from manufacturer to manufacturer. To solve these problems, a quantitative immunofluorescence (QIF) method which uses a calibrated photometric system and incorporates reducing agents into the mounting medium to reduce fading was developed to replace the visual method of endpoint determination. A uranyl glass slide was used to calibrate the instrument's voltage measurements, permitting daily comparisons and measurement of the instrument reading fluctuations. The QIF method was initially tailored to the determination of serum antibodies to Toxoplasma gondii by measuring the fluorescence intensity of individual tagged organisms. The nonspecific fluorescence intensity resulting from the counterstain was eliminated by use of a red-suppressing filter. The dilution-correlated polar fluorescence component was removed by subtraction of the intensity for the matching negative control dilution from each sample dilution intensity. The QIF method showed a 94% correlation with the visual comparison method for 62 clinical specimens.
Similar articles
-
Application of quantitative immunofluorescence to clinical serology: antibody levels of Treponema pallidum.J Clin Microbiol. 1992 May;30(5):1294-6. doi: 10.1128/jcm.30.5.1294-1296.1992. J Clin Microbiol. 1992. PMID: 1583134 Free PMC article.
-
Characterization of instrumentation and calibrators for quantitative microfluorometry for immunofluorescence tests.J Clin Microbiol. 1989 Mar;27(3):442-7. doi: 10.1128/jcm.27.3.442-447.1989. J Clin Microbiol. 1989. PMID: 2654179 Free PMC article.
-
Evaluation of a rapid screening immunoassay for antibodies to Toxoplasma gondii.J Clin Microbiol. 1987 Nov;25(11):2207-8. doi: 10.1128/jcm.25.11.2207-2208.1987. J Clin Microbiol. 1987. PMID: 3320086 Free PMC article.
-
Fc receptors on the surface of Toxoplasma gondii trophozoites: a confounding factor in testing for anti-Toxoplasma antibodies by indirect immunofluorescence.J Clin Microbiol. 1989 May;27(5):959-61. doi: 10.1128/jcm.27.5.959-961.1989. J Clin Microbiol. 1989. PMID: 2663919 Free PMC article.
-
Comparison of the Vitek Immunodiagnostic Assay System with an indirect immunoassay (Toxostat Test Kit) for detection of immunoglobulin G antibodies to Toxoplasma gondii in clinical specimens.J Clin Microbiol. 1991 Dec;29(12):2763-7. doi: 10.1128/jcm.29.12.2763-2767.1991. J Clin Microbiol. 1991. PMID: 1757547 Free PMC article.
Cited by
-
Application of quantitative immunofluorescence to clinical serology: antibody levels of Treponema pallidum.J Clin Microbiol. 1992 May;30(5):1294-6. doi: 10.1128/jcm.30.5.1294-1296.1992. J Clin Microbiol. 1992. PMID: 1583134 Free PMC article.
References
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources