Design and Performance of a Novel Interface for Combined Matrix-Assisted Laser Desorption Ionization at Elevated Pressure and Electrospray Ionization with Orbitrap Mass Spectrometry
- PMID: 28613836
- DOI: 10.1021/acs.analchem.7b01168
Design and Performance of a Novel Interface for Combined Matrix-Assisted Laser Desorption Ionization at Elevated Pressure and Electrospray Ionization with Orbitrap Mass Spectrometry
Abstract
Matrix-Assisted Laser Desorption Ionization, MALDI, has been increasingly used in a variety of biomedical applications, including tissue imaging of clinical tissue samples, and in drug discovery and development. These studies strongly depend on the performance of the analytical instrumentation and would drastically benefit from improved sensitivity, reproducibility, and mass/spatial resolution. In this work, we report on a novel combined MALDI/ESI interface, which was coupled to different Orbitrap mass spectrometers (Elite and Q Exactive Plus) and extensively characterized with peptide and protein standards, and in tissue imaging experiments. In our approach, MALDI is performed in the elevated pressure regime (5-8 Torr) at a spatial resolution of 15-30 μm, while ESI-generated ions are injected orthogonally to the interface axis. We have found that introduction of the MALDI-generated ions into an electrodynamic dual-funnel interface results in increased sensitivity characterized by a limit of detection of ∼400 zmol, while providing a mass measurement accuracy of 1 ppm and a mass resolving power of 120 000 in analysis of protein digests. In tissue imaging experiments, the MALDI/ESI interface has been employed in experiments with rat brain sections and was shown to be capable of visualizing and spatially characterizing very low abundance analytes separated only by 20 mDa. Comparison of imaging data has revealed excellent agreement between the MALDI and histological images.
Similar articles
-
A high resolution atmospheric pressure matrix-assisted laser desorption/ionization-quadrupole-orbitrap MS platform enables in situ analysis of biomolecules by multi-mode ionization and acquisition.Anal Chim Acta. 2018 May 12;1007:16-25. doi: 10.1016/j.aca.2017.12.045. Epub 2018 Jan 6. Anal Chim Acta. 2018. PMID: 29405984 Free PMC article.
-
Dual-source mass spectrometer with MALDI-LIT-ESI configuration.J Proteome Res. 2007 Feb;6(2):837-45. doi: 10.1021/pr060514i. J Proteome Res. 2007. PMID: 17269740
-
A comprehensive high-resolution mass spectrometry approach for characterization of metabolites by combination of ambient ionization, chromatography and imaging methods.Rapid Commun Mass Spectrom. 2014 Aug 30;28(16):1779-91. doi: 10.1002/rcm.6960. Rapid Commun Mass Spectrom. 2014. PMID: 25559448
-
Surface analysis of lipids by mass spectrometry: more than just imaging.Prog Lipid Res. 2013 Oct;52(4):329-53. doi: 10.1016/j.plipres.2013.04.005. Epub 2013 Apr 24. Prog Lipid Res. 2013. PMID: 23623802 Review.
-
The synergy of elemental and biomolecular mass spectrometry: new analytical strategies in life sciences.Chem Soc Rev. 2009 Jul;38(7):1969-83. doi: 10.1039/b618635c. Epub 2009 Apr 16. Chem Soc Rev. 2009. PMID: 19551177 Review.
Cited by
-
193 nm Ultraviolet Photodissociation for the Characterization of Singly Charged Proteoforms Generated by MALDI.J Am Soc Mass Spectrom. 2023 Feb 1;34(2):328-332. doi: 10.1021/jasms.2c00302. Epub 2023 Jan 9. J Am Soc Mass Spectrom. 2023. PMID: 36622763 Free PMC article.
-
Maintenance of Deep Lung Architecture and Automated Airway Segmentation for 3D Mass Spectrometry Imaging.Sci Rep. 2019 Dec 27;9(1):20160. doi: 10.1038/s41598-019-56364-4. Sci Rep. 2019. PMID: 31882724 Free PMC article.
-
External Data Systems Enable Enhanced (and Sustainable) Fourier Transform Mass Spectrometry Imaging for Legacy Hybrid Linear Ion Trap-Orbitrap Platforms.J Am Soc Mass Spectrom. 2024 Nov 6;35(11):2690-2698. doi: 10.1021/jasms.4c00145. Epub 2024 Jul 20. J Am Soc Mass Spectrom. 2024. PMID: 39031087 Free PMC article.
-
Mass Spectrometry Imaging as a Tool to Investigate Region Specific Lipid Alterations in Symptomatic Human Carotid Atherosclerotic Plaques.Metabolites. 2021 Apr 18;11(4):250. doi: 10.3390/metabo11040250. Metabolites. 2021. PMID: 33919525 Free PMC article.
-
Validation of MALDI-MS imaging data of selected membrane lipids in murine brain with and without laser postionization by quantitative nano-HPLC-MS using laser microdissection.Anal Bioanal Chem. 2020 Oct;412(25):6875-6886. doi: 10.1007/s00216-020-02818-y. Epub 2020 Jul 25. Anal Bioanal Chem. 2020. PMID: 32712813 Free PMC article.
Publication types
LinkOut - more resources
Full Text Sources
Other Literature Sources